Logger Script
본문바로가기

제품소개

Best Global Partnership NEMO TECH CO.,LTD

제품 품목별

  • homeHOME
  • >
  • 제품소개
  • >
  • 제품 품목별

상용전압 측정이 가능한 절연 다기능 기록계

제조사
GRAPHTEC
제품명
midi LOGGER GL900
제품번호
GL900

견적문의 카달로그

PRODUCT INFO. 제품상세정보

GL900-4 : Easy-to-use upright high-speed isolated 4-channel multifunction logger

GL900-8 : Easy-to-use upright high-speed isolated 8-channel multifunction logger

An easy-to-use upright device enabling isolated 8-channel multifunction input, the GL900 is capable of performing high-speed simultaneous measurements of voltage, temperature, and various other phenomena.

Voltage +/-20 mV to +/-500 V

Temperature Thermocouples:
K, J, E, T, R, S, B, N, W
Humidity 0 to 100%
(the B-530 option is required)
Pulse 4 channels
Count, Inst., Rpm
Logic 4 channels
‡ Select either Pulse or Logic

*) The input signal needs to be not exceeding the specified voltage at any time.
The component in the amplifire of the unit may have the damage by over voltage.
Please check the details of over voltage in the specification sheet (Maximum input voltage or Withstand voltage seccion) or the caution label attached on the unit.

Data can be captured to PC-friendly USB memory sticks

Long-term data can be captured directly to built-in 256-MB flash memory or to an external USB memory stick at sampling intervals of from 1 ms to 1 min. For high-speed sampling at intervals faster than 1 ms, up to one million data points can be captured to internal RAM.

Example of 8-channel analog measurement

Capture destination

10μs

100μs

500μs

1ms

10ms

100ms

1s

Internal RAM
(up to one million points)

10 seconds

Approx. 1 min. and 40 sec.

Approx. 8 min. and 20 sec.

Approx. 16 min. and 40 sec.

Approx. 2 hrs. and 40 sec.

Approx. 1 day and 3 hrs.

Approx. 11 days and 13 hrs.

Internal flash memory (256 MB)

×

×

×

Approx. 1 hour.

Approx. 11 hrs.

Approx. 4 days

Approx. 46 days

External USB memory stick (512 MB)

×

×

×

Approx. 22 hrs.

Approx. 22 hrs.

Approx. 9 days

Approx. 93 days

The USB memory stick must be a standard model (without fingerprint recognition or other proprietary features).

Can be used as an X-Y recorder

The GL900 reproduces analog X-Y recorder movements and provides the illusion of pen up/pen down movements. It can be operated like an analog X-Y recorder and can also be used as a 4-pen X-Y recorder. The digital data format facilitates post-measurement confirmation of data values and report creation.

High-precision temperature measurement even during high-speed sampling

Lets users perform high-precision temperature measurements even during high-speed sampling – ideal for performing combined voltage and temperature measurements.

Comprehensive built-in trigger and timer functions

Using a combination of trigger and timer functions eliminates superfluous data and enables capture of only the required data.

Setting example 1

To perform measurement over a four-day period starting January 10

Timer setting

Date and time

Start setting

January 10 00 hours 00 minutes

Stop setting

January 14 23 hours 59 minutes

Trigger setting

Start trigger

Off

Stop trigger

Off

Setting example 2

To perform measurements of abnormal signals during device operations

Timer setting

Daily cycle

Start setting

09 hours 00 minutes

Stop setting

17 hours 00 minutes

Trigger setting

Start trigger

Level CH 1 (3 V Rising)

Stop trigger

Level CH 1 (2 V Falling)

Repeat

On

Setting example 3

To perform measurements every 20 minutes

Timer setting

Hourly cycle

Start setting

00 minutes 00 seconds

Stop setting

20 minutes 00 seconds

Trigger setting

Start trigger

Off

Stop trigger

Off

Setting example 4

To perform measurements for a period of one hour, every four hours, daily

With the timer set to daily cycle status, data is captured repeatedly for one hour every four hours.

Trigger setting

Start trigger

Off

Stop trigger

Scheduled time (one hour)

Repeat

On (Repeat interval: 4 hours)

Timer settings

Timer mode

Off, Date and time, Daily cycle, Hourly cycle

Trigger settings

Start source setting

Off, Level value, External input

Stop source setting

Off, Level value, External input, Scheduled time

Pre-trigger

0-100%

Repeat capture

On, Off and Repeat interval

High-voltage measurement capability

The wide 500 V range enables 100 to 240 VAC power supply voltage waveform measurements. Using logic input and a clamp meter simultaneously allows measurement of a device’s power supply voltage and current concurrently with sequential control of various points.

Built-in, large-format 5.7-inch color LCD for easy-to-read waveforms

The bright, easy-to-read large-format 5.7-inch color TFT LCD provides vivid, easy-to-read waveform displays. Cursor keys enable fast, easy control and setup. The waveform display can be scrolled at high-speed – 10 ms/DIV.

Free Running display for waveform-checking without the need for data capture

The Free Running display lets users check input signal waveforms even before measurements begin. Since waveforms are displayed on each setup screen, users can make settings while viewing the waveforms.

Easy PC measurement via USB; remote monitoring via Ethernet web server and FTP functions

The USB and Ethernet connections enable transfer of captured data to your PC and setup and control of the GL900 from a PC, even without the PC software provided standard with the GL900.

Web server/FTP server functions

Waveform display and GL900 setup operations can be performed via a web browser (e.g., Internet Explorer). In addition, data files captured to the GL900’s internal memory or to a USB memory stick can be transferred or deleted from the PC.

USB drive mode

When your GL900 is connected to your PC via the USB interface, the GL900 can be operated in USB mode to enable fast, easy data transfers from internal memory to the PC.

NTP client function

Simply connect the GL900 to an NTP server via an Ethernet connection to synchronize GL900 time with NTP server time at periodic intervals

Dedicated software for real-time data capture

Three measurement screens are provided to allow selection of the screen that best suits measurement needs.
The Replay screen provides a Zoom screen feature to enable enlarged display of specific sections of long-term measurement data.

Simple operations for anyone

Easy-to-use software using icon keys for intuitive operations

Convenient functions

Various convenient data-processing functions are built in.

  • Direct to Excel function
    This function enables measurement data to be written directly to an Excel file.
  • Search function
    This function enables searching for specific values in the captured data.
  • CSV batch conversion function
    This function enables batch conversion of multiple captured files to CSV file format.
  • Thumbnail function
    This function enables display of captured data files as thumbnails.


PRODUCT INFO. 제품상세정보

GL900 main unit specifications


Item

Description

GL900-8

GL900-4

No. of analog input ch.

8 ch

4 ch

External input/output

Trigger input (1 channel), Logic input (4 channels) or Pulse input (4 channels), Alarm output (4 channels)

Sampling interval

10μs to 1 min

TIME/DIV

10 ms/DIV to 24 hour/DIV

Timer functions

Date and time, daily cycle, hourly cycle

Trigger functions

Type

Start: Data capture starts when a trigger is activated; Stop: Data capture stops when a trigger is activated

Condition

Start: Off, Input signal level (analog, logic/pulse), External*1, Scheduled time

Stop: Off, Input signal level (analog, logic/pulse), External *1, Scheduled time

Combination

Input signal level: Level OR, Level AND, Edge OR, Edge AND

Mode

H (Rising), L (Falling), Window In*2, Window Out*2

Alarm setting functions

Rising, Falling, Window In*2, Window Out*2

Alarm output *1

Number of channels: 4, Open collector output (5V, 10 kΩ pull-up resistance)

Pulse input*1, *3

RPM mode

50 to 20 M RPM/F.S. (in steps of 1, 2, or 5)

Count mode

50 to 20 M C/F.S. (in steps of 1, 2, or 5)

Inst. Mode

50 to 20 M C/F.S. (in steps of 1, 2, or 5)

Calculation functions

Statistical calculations *4: Average, Peak, Maximum, Minimum, RMS (2 calculations can be set simultaneously)

Other functions

Search function, annotation input function

PC inteface

Ethernet (10BASE-T/100BASE-TX), USB (High Speed supported) provided as standard

Ethernet functions

Web server function, FTP server function, NTP client function

USB function

USB drive mode (File transfer and deletion from internal GL900 memory)

Memory device

Internal

One million data points / Internal flash memory:Approx. 256 MB

External

USB memory slot (High speed supported) *5

Display screens

Waveforms + digital values, enlarged waveforms, digital values + calculation results, X-Y

Display unit

5.7-inch TFT color LCD

Operating environment

0 to 40℃, 5 to 85% R.H. (15 to 35℃ when using batteries)

Withstand voltage

Between each input channel and GND: 1000 V p-p for one minute, between input terminals: 1000 Vp-p for one minute

Power supply

AC adapter

100 to 240 VAC, 50 to 60 Hz

DC input

8.5 to 24 VDC

Battery pack *6

Option

Power consumption

42VA (when operating and charging battery with AC power)

40VA (when operating and charging battery with AC power)

External dimensions

232 x 150 x 80 mm (W x H x D), approx.

Weight (approx.)

1.1 kg (excluding AC adapter and battery)

1.0 kg (excluding AC adapter and battery)

Vibration-tested conditions

Equivalent to automobile parts Type 1 Category A classification


Terminal block specifications


Item

Description

GL900-8

GL900-4

Number of input channels

Fixed to 8 channels

Fixed to 4 channels

Input terminal type

Voltage

BNC connector

Temperature

M3 screw type terminal board *7

Input method

All channels isolated Imbalanced input Simultaneous sampling of all channels

Measurement ranges

Voltage

20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100, 200, 500 V F.S., 1-5 V F.S.

Temperature

Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26)

Humidity

0 to 100% (voltage 0 V to 1 V scaling conversion) * with B-530 (option)

Input filter

Off, Line, 5 Hz, 50 Hz, 500 Hz

Measurement accuracy *8
(23°C±5°C)
When 30 minutes or more have elapsed after power was switched on Filter : Line
GND : connected

Voltage

±0.25% of F.S.

Thermocouple

Type

Measurement temperature range

Measurement accuracy

R/S

0°C ≦ TS ≦ 100°C

±7.0°C

100°C < TS ≦ 300°C

±5.0°C

R:300°C < TS ≦ 1600°C

±(0.05% of rdg +3.0°C)

S:300°C < TS ≦ 1760°C

±(0.05% of rdg +3.0°C)

B

400°C ≦ TS ≦ 600°C

±5.5°C

600°C < TS ≦ 1820°C

±(0.05% of rdg +3.0°C)

K

-200°C ≦ TS ≦ -100°C

±(0.05% of rdg +3.0°C)

-100°C < TS ≦ 1370°C

±(0.05% of rdg +2.0°C)

E

-200°C ≦ TS ≦ -100°C

±(0.05% of rdg +3.0°C)

-100°C < TS ≦ 800°C

±(0.05% of rdg +2.0°C)

T

-200°C ≦ TS ≦ -100°C

±(0.1% of rdg +2.5°C)

-100°C < TS ≦ 400°C

±(0.1% of rdg +1.5°C)

J

-200°C ≦ TS ≦ -100°C

±3.7°C

-100°C < TS ≦ 100°C

±2.7°C

100°C < TS ≦ 1100°C

±(0.05% of rdg +2.0°C)

N

0°C ≦ TS ≦ 1300°C

±(0.1% of rdg +2.0°C)

W

0°C ≦ TS ≦ 2000°C

±(0.1% of rdg +2.5°C)

Reference contact compensation accuracy : ±1.0°C
‡ Thermocouple diameters T: 0.32mm, others: 0.65mm

A/D converter

16 bits (out of which 14 bits are internally acknowledged)

Input resistance

1MΩ ±5%

Allowable signal source resistance

Up to 1kΩ

Maximum permissible input voltage

Between + and - input terminals

20 mv to 1 V : ±30VDC
2 V to 500 V : ±500VDC

Between input channel terminals

60 Vp-p

Between input channel terminal and GND terminal

60 Vp-p

Withstand voltage

Between input channel terminal and GND terminal

1 minute at 1000 Vp-p

Between input channel terminals

1 minute at 1000 Vp-p


*1
Logic alarm cable (B-513) is required.
*2
Cannot be set for logic input
*3
Maximum input frequency: 50 kHz, maximum number of counts: 15 M C
*4
In real time or when Between Cursors has been specified (during Replay)
*5
1 file = 2 Gbytes (depends on the USB memory stick used)
*6
Please install two battery packs.
*7
Connections are made to both the BNC terminal and M3 screw terminal for the same channel.
*8
Thermocouple diameters T:0.32φ, others:0.65φ
*9
Operating temperature range: -25 to +80°C


Control software specifications


Item

Description

Supported OS

Windows 2000, Windows XP, Windows Vista (32-bit and 64-bit versions)

Functions

GL900 control, real-time data capture, data conversion

Setting range

Amp settings, data capture settings, trigger settings, alarm settings, other

Captured data

Real-time data

Binary: Sampling speed: 10 μs to 60 s
CSV: Sampling speed: 10 ms to 60 s

Data conversion

Binary, CSV

Display information

Analog waveforms, logic waveforms, pulse waveforms, digital values

File conversion

Data between cursors, All data

2-screen function (Zoom)

Display of current and past data

Display of statistics and history

Display of maximum, minimum, and average values


PRODUCT INFO. 제품상세정보

Options and accessories


Product name

Model name

Specification

Battery pack*6

B-517

One pack

Logic alarm cable

B-513

2 m

DC drive cable

B-514

2 m

Humidity sensor*9

B-530

3 m

Safe probe

RIC-141

1:1, 42 pF

BNC-BNC cable

RIC-112

1.5 m

BNC banana plug cable

RIC-113

1.5 m

BNC alligator clip cable

RIC-114

1.5 m

Rod-shaped K-type thermocouple

RIC-410

1.1 m

K-type thermocouple for static surfaces

RIC-420

1.1 m

L-shaped K-type thermocouple for static surfaces

RIC-430

1.1 m


Battery pack
(B-517)
Logic alarm cable
(B-513)
DC drive cable
(B-514)
Humidity sensor
(B-530)
Safe probe
(RIC-141)
Rod-shaped K-type
thermocouple
(RIC-410)
K-type thermocouple
for static surfaces
(RIC-420)
L-shaped K-type thermocouple
for static surfaces
(RIC-430)


목록